USER INTERFACE
2.13.2018
99
13.3.2. out commands
out commands:
Setting parameters or temperature values.
Command
Parameter
Response of instrument
out_mode_01
0
Use working temperature (Setpoint 1).
out_mode_01
1
Use working temperature (Setpoint 2).
out_mode_01
2
Use working temperature (Setpoint 3).
out_mode_02
0
Selftuning “off”.
Temperature control using the stored parameters.
out_mode_02
1
Selftuning “once”
Single selftuning of the controlled system after the next start.
out_mode_02
2
Selftuning “always”
Continual selftuning of the controlled system whenever a new setpoint is
to be reached.
out_mode_03
0
Set external programmer input to voltage. Voltage 0 V … 10 V
out_mode_03
1
Set external programmer input to current. Current 0 mA … 20 mA
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_mode_08
0
Set the control dynamics – aperiodic
1
out_mode_08
1
Set the control dynamics - standard
out_mode_09
0
Set pump to stage control.
out_mode_09
1
Set pump to pressure control.
out_mode_09
2
Set pump to flow rate control.
out_mode_09
3
Set pump to flow rate via the external bypass.
out_sp_00
xxx.x
Set working temperature. (Setpoint 1).
out_sp_01
xxx.xx
Set working temperature. (Setpoint 2).
out_sp_02
xxx.xx
Set working temperature. (Setpoint 3).
out_sp_03
xxx.x
Set upper temperature limit.
out_sp_04
xxx.x
Set lower temperature limit.
out_sp_06
xxx.xx
Set watchdog set point.
1
The dynamics cannot be changed with the maximum temperature gradient, heating/cooling active.