5.1.3. "Adjust control performance" menu
Select menu
indicates available submenu.
Preparing for external control:
Connect the Pt100 external sensor to the socket in the socket
panel. It is normally not necessary to calibrate the sensor.
In special situations, a three-point calibration, for example, can be
performed with the "Adjust sensors" function (see page 83).
The buttons will display the current
settings.
Select parameters
Set new
values.
Internal parameters
External parameters
or
Internal/external control parameters
In most cases, the factory-set control parameters will be
adequate for achieving an optimal temperature sequence in
the item being controlled.
Adjustable control parameters give you the ability to adapt to
unusual processes.
Setting range:
internal/external
0.1 ... 99.9 K
Proportional range >Xp<
The proportional range is the temperature range below the
setpoint in which the heating capacity is controlled from 100 %
to 0 %.
Setting range:
internal/external
0 ...10000 s
Reset time >Tn<
(integral proportion)
Compensation for the control deviation that remains due to the
proportional controller. Reset times that are too small may lead
to instability. Reset times that are too large will make
compensation of the control difference unnecessarily long.
Setting range:
internal/external
0 ... 1000 s
Rate time >Tv<
(differential proportion)
The differential proportion shortens the adjustment time.
If the rate time is too small, equalization of an interference
value will be extended and you will experience large
overshoots when approaching a setpoint. If the rate times are
too great, you may experience instability (oscillations).
Setting range:
0.1 ... 99.9 K
Proportional range >Xpu<
The Xpu proportional range of the underlying controller is
needed only for external control.