"Connect unit" menu
58
2.13.2018
Timeout
The timeout time in seconds can be set here. The watchdog must
be reset within the timeout time, otherwise, a timeout is triggered.
Setpoint value
The safety setpoint value can be set here. If a timeout occurs and
the watchdog mode is set to “Setpoint”, the device adjusts to this
setpoint
Restart mode:
Here, the commands can be set that trigger a watchdog reset.
(They are thus used to prevent timeouts).
The following options can be selected:
All commands:
Every received valid interface command
resets the watchdog.
Setpoint command:
Only the “Setpoint” command
(OUT_SP_00) resets the watchdog.
Create Warning Signal:
Here it can be set, whether a warning should be shown on the
display, when a timeout occurs.
The following options can be selected:
Off:
The warning is not shown on the
display.
On:
The warning is shown on the display.
Reset warning:
Here it can be set, whether the warning displayed due to a timeout
occurrence should be reset automatically on receipt of a new valid
command.
The following options can be selected:
Manual:
The user must acknowledge the
warning manually on the display.
Auto:
The warning is reset automatically if a
valid command is received.