"Install unit" menu
84
2.13.2018
Select
>Status< >Inactiv<
.
Select
>
Edit
<
Example: Calibration of the
external
sensor.
Important:
During calibration
>Status< >inactive<
must be activated.
Switch to
>activ<
after calibration.
Select >
Number of points
<.
Example:
>3-point< calibration
This setting determines the
number of the following steps.
Follow the instructions on the
screen.
You can perform a
>1-point<, >2-point<,
or
>3-point< calibration
(Values are examples only).
Calibration point 1
Set >
Setpoint
<
Wait
Set >
Calibration value
<
It is not possible to enter a
calibration value while a button is
light grey.
Switch window
Wait