Error messages
116
2.13.2018
Alarm-Code
Cause
Diagnosis / Remedy
E1210
Error in current measurement for
compressor stage 1.
If this error continues to occur, please
contact JULABO Service.
E1211
Error in current measurement for
compressor stage 2.
If this error continues to occur, please
contact JULABO Service.
E1212
Error in current measurement for
compressor stage 2.
If this error continues to occur, please
contact JULABO Service.
E1212
Error in current measurement in heater
circuit.
If this error continues to occur, please
contact JULABO Service.
E1213
Error in current measurement of the pump.
If this error continues to occur, please
contact JULABO Service.
E1214
Error in current measurement of the pump.
If this error continues to occur, please
contact JULABO Service.
E1215
Error in main power voltage measurement.
If this error continues to occur, please
contact JULABO Service.
E1216
Error in main power voltage measurement.
If this error continues to occur, please
contact JULABO Service.
E1216
Error in voltage measurement in heater
circuit.
If this error continues to occur, please
contact JULABO Service.
E1217
Error in main power frequency
measurement.
If this error continues to occur, please
contact JULABO Service.
E1218
Error in main power frequency
measurement.
If this error continues to occur, please
contact JULABO Service.
E1301
The temperature at the power module is
higher than the configured limit.
Turn the unit off and ensure sufficient
cooling.
E1302
The heating is locked, since the pump
pressure is too low.
Check the settings for the pump.
E1303
The cooling for the electronics is too low.
If this error continues to occur, please
contact JULABO Service.
0
E1421
Ambient temperature out of specifications.
Please check the ambient temperature and
clean the condenser.
E1422
Ambient temperature out of specifications.
Please check the ambient temperature and
clean the condenser.
E1423
Compressor temperature too low.
Please check the ambient temperature.
E1424
Compressor temperature too high.
Please check the ambient temperature and
clean the condenser.
0
E1427
Error in stage 1 of the cooling system.
One or more of the following problems has
occurred:
(1) Ambient temperature too high
(2) Condenser contaminated
(3) Cooling water temperature too high
(4) Quantity of cooling water too low