Normal display
8
2.13.2018
Warning
•
The >
TANK
< high temperature cut-off should be set to
15 °C above the working temperature setpoint.
•
The >
RESERVOIR
< high temperature cut-off
must
be set
to at least 25 °C below the heattransfer liquid's flashpoint.
Adjust the normal display
You can adjust the normal
display to your requirements
by pressing the >Display< key.
You may select the colors
used in the normal display's
chart.
>
Settings menu
< page 18.
Chart shows the progression of the setpoint and actual
temperature.
Displays during errors
ALARM
red
>
WARNING
yellow
>
Help is always accessible through
the icons
or
.
Touch the icon and a list of errors
will be displayed.
The unit provides straightforward and intuitive operation on the
color TFT display even during errors.
Error messages are divided into two categories:
>ALARM< and >WARNING<
Date and time when the error appeared are stored and displayed.
If possible, this data will also be stored during removal of the error.
Example code