USER INTERFACE
2.13.2018
47
For setting range see display of unit
A High temperature cut-off (tank)
B Upper temperature limit
C Lower temperature limit
D Setpoint
The warning function will be activated only when the temperature
value is within the selected limit values for three seconds after
starting from the "OFF" condition.
9.1.3. Allowed temperature gradient per minute
Maximum allowed
temperature gradient
per minute during
heating
up /
cooling
down.
When the gradient limiter is activated, a setpoint step is executed
as a gradient with the values selected here.
This feature is valuable e.g. to safely implement temperature
changes with a sensitive glass reactor.
•
Activated limitation is shown as ticker on the regular
display.
Select
and set value
For setting range see display of unit
The buttons will display the current
settings
Choose button and edit value
or deactivate the function.
Attention:
The setting range for >
Gradient
<, in the chapter of programmer, is
limited by these values (page 36).