USER INTERFACE
2.13.2018
31
Notice
Settings of setpoint limits for
>
Pumpstage
< and
>
Limit Pressure
< are active.
Examples:
If the >Limit Pressure< is exceeded at >stage 3< an alarm including
the cut-off of the unit is activated!
5.2.1.2.
Type „Pressure control“
Select .
Set pressure setpoint.
Example „0.5“
Setpoint limits
refer to page 48
The setting ranges depend on the performance category of the
temperature control system.
Example: see display of unit.
Settings in the >
" Adjust safety " menu
< will influence these
values. If a >
Setpoint limit
< is set, you will not be able to
exceed or fall below this value, respectively.
You will receive a message stating "Value is too small or too
large"
Example:
The displayed value, in this case „Max:0.65“, always refers to
the next higher limit.