2.13.2018
Alarm-Code
Cause
Diagnosis / Remedy
E61
CAN bus error.
The data communication between the TFT
and the refrigeration module is permanently
monitored. If communication cannot be
established (e.g. by a defective connection
cable), the error message „E 61“ is
generated. By pressing the „OK“ key, the
alarm is acknowledged. The device
continues operating purely as heating
circulator until the next interruption of the
power supply (power-off).
When the fault has been repaired, the TFT
controls the refrigeration module according
to the settings in the menu after the next
power-on.
E62
CAN bus error.
Turn off the unit at the power switch, wait 4
seconds, and turn the unit back on.
E63
Error in the electronics.
Turn off the unit at the power switch, wait 4
seconds, and turn the unit back on.
E67
Error in module configurations
If this error continues to occur, please
contact JULABO Service
E102
The line to an internal temperature sensor is
interrupted.
If this error continues to occur, please
contact JULABO Service.
E104
Error in internal temperature measurement
circuit
If this error continues to occur, please
contact JULABO Service.
E105
The line to an internal temperature sensor is
short-circuited.
If this error continues to occur, please
contact JULABO Service.
E106
Error in internal temperature measurement
circuit.
If this error continues to occur, please
contact JULABO Service.
E107
Error in internal temperature measurement
circuit.
If this error continues to occur, please
contact JULABO Service.
E108
The relay lock of the safety mechanism is
still active.
The unit was turned back on too quickly to
reset the relay lock of the safety
mechanism.
Turn off the unit at the power switch, wait 4
seconds, and turn the unit back on.
E110
The line to an internal temperature sensor is
interrupted.
If this error continues to occur, please
contact JULABO Service.
E112
Error in internal temperature measurement
circuit.
If this error continues to occur, please
contact JULABO Service.
E113
The line to an internal temperature sensor is
short-circuited.
If this error continues to occur, please
contact JULABO Service.
E114
Error in internal temperature measurement
circuit.
If this error continues to occur, please
contact JULABO Service.
E115
Error in internal temperature measurement
circuit.
If this error continues to occur, please
contact JULABO Service.