"Determine thermodynamics" menu
26
2.13.2018
A Standard
B Aperiodic
C Temperature stability
D Setpoint
E Temperature ramp
>
Dynamic <
This parameter influences the temperature sequence only
during
internal
control.
Available parameters:
Standard
The temperature will climb faster, but may overshoot
by up to 5%. If a ramp is defined, the temperature sequence
will largely follow this ramp.
Aperiodic
.
Temperature will increase with time offset (no
overshoots).
Both settings will achieve adequate temperature stability after
approximately the same amount of time.
Setting range: 0.00 to 5.00
S
Setpoint
Ext External temperature
Int
Internal temperature
>CoSpeed factor<
This parameter will influence the temperature sequence only
with
external
control.
The setting influences calculation of the control parameters
during identification, thereby influencing control behavior.