Remote control
100
2.13.2018
Command
Parameter
Response of instrument
out_sp_07
x
Set the pump pressure stage. (1 ... 4).
[not on A30]
out_sp_08
xxx.xx
Set flow rate setpoint.
[not on A30]
out_sp_09
xxx.xx
Set pump pressure setpoint.
[not on A30]
out_sp_10
xxx
Set variable via the serial interface -100 … 100 [%]
out_sp_11
0
Temperature display in
°C
.
out_sp_11
1
Temperature display in
°F
.
out_sp_12
0
Pressure display in
bar
.
out_sp_12
1
Pressure display in
psi
.
out_sp_13
0
Flow rate display in
l/min
.
out_sp_13
1
Flow rate display in
gpm
.
out_sp_14
xxx.xx
Set pressure warning limit, upper.
out _sp_15
xxx.xx
Set pressure warning limit, lower.
out _sp_16
xxx.xx
Set pressure alarm limit (5 s).
out _sp_17
xxx.xx
Set pressure alarm limit (1 s).
out _sp_18
xxx.xx
Set flow rate warning limit, upper.
out _sp_19
xxx.xx
Set flow rate warning limit, lower.
out_sp_25
xxx.xx
adjust maximum temperature gradient, heating
-
deactivate maximum temperature gradient, heating
out_sp_26
xxx.xx
adjust maximum temperature gradient, cooling
-
deactivate maximum temperature gradient, cooling
out_hil_00
-xxx
Set the desired maximum cooling power (0 % to 100 %).
Note:
Enter the value with a preceding negative sign!
out_hil_01
xxx
Set the desired maximum heating power (10 % to 100 %).
out_par_04
xxx
Control parameter CoSpeed of the external
controller 0 ... 5.0.
out_par_06
xxx
Xp control parameter of the internal controller.
out_par_07
xxx
Tn control parameter of the internal controller.
out_par_08
xxx
Tv control parameter of the internal controller.
out_par_09
xxx
Xp control parameter of the cascade controller.
2
If the maximum temperature gradient, heating/cooling, values change, the dynamics of the internal control is
automatically adjusted to “Standard”