USER INTERFACE
2.13.2018
101
Command
Parameter
Response of instrument
out_par_10
xxx
Proportional portion of the cascade controller.
out_par_11
xxx
Tn control parameter of the cascade controller.
out_par_12
xxx
Tv control parameter of the cascade controller.
out _par_13
xxx
Maximum internal temperature of the cascade controller.
out _par_14
xxx
Minimum internal temperature of the cascade controller.
out_par_15
xxx
Upper band limit 0 ... 200 K.
out_par_16
xxx
Lower band limit 0 ... 200 K.
13.4. Status messages
Status messages
Description
00 MANUAL STOP
in "OFF" state.
01 MANUAL START
in normal control mode.
02 REMOTE STOP
in "Remote OFF" state.
03 REMOTE START
in remote control mode.
13.4.1. Status messages as reply to sent commands
Messages
Description
-08 INVALID COMMAND
The unit did not recognize the most recently received command.
-09 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
The most recently received command is not permitted in the current
operating mode (example: setpoint specification while the unit is working
in manual mode).
-10 VALUE TOO SMALL
Entered value too small.
-11 VALUE TOO LARGE
Entered value too large.
-13 VALUE EXCEEDS
TEMPERATURE LIMITS
Value lies outside the adjusted range for the high and low temperature
warning limits. But the value is stored.
Note:
In addition to status messages, error messages are also transferred
(See error messages starting on Page 108).