"Determine thermodynamics" menu
28
2.13.2018
5.1.4. "Adjust limits" menu
Select menu
The buttons will display the current
settings.
Example: setting ranges see
display of unit.
The >Limits< menu allows you to define the minimum and
maximum values for all important setting ranges and power
variables.
The setting ranges depend on the performance category of the
temperature control system.
Selected maximum heating / cooling capacity
The unit's heating and cooling capacities are adjustable.
100% corresponds to the capacity specified in the technical
data.
Setting range:
Max. heating capacity 0 to 100% in 1% steps
Max. cooling capacity 0 to 100% in 1% steps
Min. internal temp
and
max. internal temp
Maximum and minimum setpoint in internal bath.
The max. internal temp and min. internal temp limits apply only
when using the "external" operating mode. Max. internal temp and
min. internal temp define static limits for the anticipated
temperatures in the internal bath. The temperature controller
cannot exceed these limits, even if this would be necessary in
order to achieve the desired temperature in the external system. In
some situations this may prevent you from reaching the external
setpoint.
Reasons for setting limits:
Protect the heat transfer liquid from overheating.
Prevent the high temperature cut-off >Error 14< from
triggering an undesired alarm shutoff.
Set the >
Internal max
.<
value at least 5 °C below the
>High temperature cut-off (tank)< value.
Protect the pump motor from excessive viscosity of the
heat transfer liquid at low temperatures.