Box stuck at input
Error
• One of the following error messages is added to Alarm Manager:
"E506: No box detected at box input.", "E500: timeout retrieving box
at input" or "E200: Timeout, reaching limit switch pusher box failed".
• Process is stopped.
• Remove box.
• Acknowledge alarm in the alarm manager.
• Initialize the system before Autosampler III can be restarted.
Box stuck at output
Warning
• Popup in HMI with message "Box output blocked."
• Remove box.
• Press "Ok" in popup.
• Autosampler III will continue.
Camera cannot
process box:
Wrong job loaded
No samples in box
Warning
• Warning is added to Alarm Manager "Product not matching recipe,
moving box out."
• Box is automatically outputted.
• Autosampler III will continue.
Mismatch between
product and
barcode
Warning
• Warning is added to Alarm Manager "Product not matching recipe,
moving box out."
• Box is automatically outputted.
• Autosampler III will continue.
No plate in stacker
Warning
• Warning is added to Alarm Manager "Detecting plate failed, please
reload plates."
• Autosampler III retries to de-stack plate.
• Popup in HMI if still no plate was found with message "Reload plates
in de-stacker".
• Reload plates in de-stacker.
• Press "Ok" in popup.
• Autosampler III will continue.
Plate barcode not
scanned
Warning
• Warning is added to Alarm Manager "Unable to scan plate barcode."
• Autosampler III retries to de-stack plate.
• Popup in HMI if still no plate was found with message "Reload plates
in de-stacker".
• Reload plates in de-stacker.
• Press "Ok" in popup.
• Autosampler III will continue.
Plate stuck at
stacker
Error
• Error is added to Alarm Manager "E200: Timeout, reaching limit
switch stacker pins failed."
• Process is stopped.
• Remove plate.
• Acknowledge alarm in the alarm manager.
• Initialize the system before Autosampler III can be restarted.
Plate pushout
sensor did not
detect plate
Error
• Error is added to Alarm Manager "E501: No plate detected at
plate sensor."
• Remove plate by opening cover.
• Press "Reset" button.
• Wait for air pressure.
• Acknowledge alarm in the alarm manager.
• Initialize the system before Autosampler III can be restarted.