www.hygiena.com
26
Step 3: Wait for the Autosampler III to start up.
After pressing the "Reset" button, the Autosampler III will need up to 45 seconds to start. The status light will be
blue when the Autosampler III's start-up is finished. During start-up, the status light will be blinking red, this
is due to the soft start of the pressure regulator. If no air pressure is available, a soft start timeout alarm will
be generated in the alarm manager of Xperimate. If a correct start-up is performed, the light will automatically
turn blue after the start-up time. If the light is still blinking red after the start-up time has passed, it indicates
a faulty start-up, possibly due to an "Emergency" button still being pressed, no air pressure available or an
unlocked door.
Step 4: Login to the Autosampler III
When the Autosampler III is turned on, the user must log into the onboard computer in order to run the
Autosampler III. The user must log into both the windows environment and the control software environment.
To log into both environments please follow the following instructions:
Windows:
•
Select the appropriate account: for day-to-day work please log in with (user account)
•
Enter the correct windows credentials.
Xperimate:
•
Double click on the Xperimate program icon on the desktop
•
Fill in the username.
•
Fill in the correct password and press "OK".
•
The user is now logged into the control panel of the Autosampler III. The image below shows the
overview presented after logging in.
First view control panel after login.