www.hygiena.com
67
Pneumatic Components Maintenance Schedule
Part
Inspect
Clean
Lubricate
Cylinder
Monthly
Monthly
If applicable monthly,
otherwise never
Air service unit
Monthly
Monthly
Monthly
Reducer filter
Min monthly, no longer
than once per 6 months;
frequency depending on
the air conditions
DO NOT clean, always
replace (5 μL)!!
Not applicable
Filter
Monthly
Removing water by
draining
Not applicable
Couplings and valves
Monthly
Once every six months Not applicable
Compressed air tubing
Monthly
Once every six months Not applicable
12. Maintenance service panel visualizations
12.1. Introduction
For maintenance and / or manual control purposes, individual components that are connected to the PLC can
be controlled from a web interface.
There are a total of 4 pages for the different components:
•
Actuators - http://192.168.0.10:8080/autosampler_act.htm
•
Anybus - http://192.168.0.10:8080/autosampler_anybus.htm
•
Motors - http://192.168.0.10:8080/autosampler_mot.htm
•
Routines - http://192.168.0.10:8080/autosampler_rou.htm
12.2. Main
Every visualization is password protected
and will contain a "Main" tab allow login
or logout.
Username: Admin
Password: prepare
12.2.1. Login
•
Press "Login" to log onto the Autosampler III.
•
Fill in "Username".
•
Fill in "Password".
•
Press "Ok" to login
Now all the controls on the other tabs will be
enabled and it will be possible to control the
Autosampler III using the different buttons.
Reserved for certified engineers.
Operations to be carried out only by certified personnel.