Chapter 9
245
Performance Verification Tests
13. Frequency Response
Figure 9-13
Frequency Response Test Setup,
≥
100 kHz
CAUTION
Use only 75
Ω
cables, connectors, or adapters on instruments with 75
Ω
connectors, or damage to the connectors will occur.
3. Connect the equipment as shown in
Figure 9-12
. Be sure to use a
mechanical adapter on analyzers with 75
Ω
inputs.
4. Press
Blue Key
,
Special 00
on the synthesized signal generator. Set
the synthesized signal generator controls as follows:
FREQUENCY
,
1 MHz
INCREMENT SET
,
50 MHz
AMPLITUDE
,
5 dBm
5. Adjust the synthesized signal generator amplitude for 0 dBm
reading on the measuring receiver.
6. Record the power meter reading in column 4 of
Table 9-15
, taking
into account the calibration factors of both the 100 kHz to 1500 MHz
power sensor and the 75
Ω
power sensor.
7. On the synthesized signal generator, press
FREQUENCY
,
50 MHz
.
8. Repeat
step 5
and
step 6
, entering each power sensor’s calibration
factor into the respective power meter.
9. On the synthesized signal generator, press
FREQUENCY
, and
↑
(up-arrow key), to step through the remaining frequencies listed in
Table 9-15
. (No measurement is made at 100 kHz).
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...