Chapter 9
231
Performance Verification Tests
9. Input Attenuation Switching Uncertainty
Table 9-12
Input Attenuation Switching Uncertainty Worksheet
Measure Switching Uncertainty
1. Set the 1 dB and 10 dB step attenuators for the total nominal
attenuation as indicated in
Table 9-12
for the 0 dB spectrum
analyzer attenuation setting.
Table 9-11
may be used as a reference
for setting the step attenuators to achieve the desired total nominal
attenuation.
2. Set the spectrum analyzer reference level as indicated in
Table 9-12
for the 0 dB spectrum analyzer attenuation setting.
For analyzers with 50
Ω
inputs, use the column labeled “(dBm)”.
For analyzers with 75
Ω
inputs, use the column labeled “(dBmV)”.
3. Set the spectrum analyzer attenuation to 0 dB.
4. Press
Single Sweep
and wait for completion of a new sweep.
5. Press
Peak Search
. Record the marker amplitude in
Table 9-12
as the
Mkr
∆
Reading.
Spectrum Analyzer
Total
Nominal
Attenuation
Attenuation
Error
Mkr
∆
Reading
TR Entry
(Switching
Error)
Attenuation
Reference Level
dB
dBm
dBmV
dB
dB
dB
dB
10
−
50
−
1.2
50
0
0
Ref
0
−
60
−
11.2
60
1
5
−
55
−
6.2
55
2
15
−45
+3.8
45
3
20
−
40
+8.8
40
4
25
−35
+13.8
35
5
30
−
30
+18.8
30
6
35
−25
+28.8
25
7
40
−
20
+23.8
20
8
45
−15
+33.8
15
9
50
−
10
+38.8
10
10
55
−
5
+43.8
5
11
60
0
+48.8
0
12
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...