300
Chapter 9
Performance Verification Tests
23. Harmonic Spurious Outputs, Options 1DN and 1DQ
RES BW, 30 kHz
LOG dB/DIV
,
10 dB
CF STEP SIZE
,
10 MHz
5. Set up the microwave spectrum analyzer by performing the
following steps:
Note that the following steps are for an HP 8566A/B microwave
spectrum analyzer, the steps may be different if you are using
another microwave spectrum analyzer.
• Press
SIGNAL TRACK
(On). Wait for the signal to be displayed at
center screen.
• Press
Mkr
→
Ref Lvl
.
• Press
Peak Search
,
SIGNAL TRACK
(Off),
MARKER
∆
• Press
CENTER FREQUENCY
and the step-up key to tune to the
second harmonic. Press
Peak Search
. Record the marker
amplitude reading in
Table 9-26
as the 2nd Harmonic Level for
the 10 MHz Tracking Generator Output Frequency.
• Perform this step only if the Tracking Generator Output
Frequency is less than 500 MHz. Press
CENTER FREQUENCY
and
the step-up key to tune to the third harmonic. Press
Peak Search
.
Record the marker amplitude reading in
Table 9-26
as the 3rd
Harmonic Level for the 10 MHz Tracking Generator Output
Frequency.
• Press
MARKER
(Off).
6. Change the spectrum analyzer center frequency to the next
frequency listed in
Table 9-26
.
7. Change the microwave spectrum analyzer center frequency and
center frequency step size to the next frequency listed in
Table 9-26
,
then repeat
step 5
. Note that the microwave spectrum analyzer
frequency is the same as the Tracking Generator Output Frequency.
Table 9-26
Harmonic Spurious Responses Worksheet
Tracking
Generator
Frequency
2nd
Harmonic
Level (dBc)
3rd
Harmonic
Level (dBc)
10 MHz
100 MHz
300 MHz
750 MHz
N/A
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...