Chapter 9
241
Performance Verification Tests
12. Absolute Amplitude Accuracy
2. Calibrate the measuring receiver and low-power power sensor.
75
Ω
Input: Calibrate the measuring receiver and 75
Ω
power sensor.
3. Connect the signal generator output to the low-power power sensor
through the Type-N cable, using an adapter.
75
Ω
Input: Connect the signal generator output to the 75
Ω
power
sensor through the Type-N cable using minimum loss pad and other
adapters as necessary.
4. Adjust the signal generator’s power level for a power meter reading
of
−
27 dBm.
75
Ω
Input: Adjust the signal generator’s power level for a power
meter reading of
−
24 dBm. Allow the power sensor adequate time to
settle; the 75
Ω
power sensor is being used on its lowest range.
Record the power meter reading here.
Power Meter Reading ____________ dBm
5. Press the following analyzer keys:
Preset, System, Alignments, Align Now, All
(wait for alignment to
finish),
Done
,
Auto Align
,
Off
.
Frequency
,
50 MHz
Span
,
2 kHz
BW/Avg
,
Resolution BW Auto Man
,
3 kHz
Video BW Auto Man 10 kHz
6. Press
Amplitude
−
25 dBm, Atten Auto Man 10 dB
,
Scale Type Log Lin
(Lin).
75
Ω
Input: Set the reference level to +26.76 dBmV.
7. Press
Det/Demod, Detector, Sample, Done
.
8. Disconnect the power sensor from the Type-N cable. Connect the
Type-N cable to the analyzer 50
Ω
input.
75
Ω
Input: Connect the Type-N cable to the analyzer 75
Ω
input
using a minimum loss adapter.
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...