Chapter 9
289
Performance Verification Tests
21. Absolute Amplitude, Vernier, and Power Sweep Accuracy, Options 1DN and 1DQ
Figure 9-24
Absolute Amplitude, Vernier, and Power Sweep Accuracy Test
Setup
CAUTION
Use only 75
Ω
cables, connectors, or adapters on instruments with 75
Ω
connectors, or damage to the connectors will occur.
2. Zero and calibrate the measuring receiver and 100 kHz to 1500 MHz
power sensor in log mode (power reads out in dBm), as described in
the measuring receiver operation manual. Enter the power sensor’s
50 MHz Cal Factor into the measuring receiver.
3. Connect the 100 kHz to 1500 MHz power sensor to the RF Out 50
Ω
as shown in
Figure 9-24
.
75
Ω
RF Out only: Connect the power sensor to the RF Out 75
Ω
as
shown in
Figure 9-24
.
4. On the spectrum analyzer, press:
Source Amptd, Amplitude
,
On Off (On), 0 dBm
(50
Ω
RF Output only)
Source Amptd, Amplitude
,
On Off (On), +42.76 dBmV
(75
Ω
RF Output only)
Single Sweep
Source Amptd
,
Attenuation Auto Man, 0 dB
5. Read the power level displayed on the measuring receiver and record
the result as TR Entry 1 of the performance verification test record
as the Absolute Amplitude Accuracy.
75
Ω
RF Out only: Add 6 dB to the power level displayed on the
measuring receiver and record the result as TR Entry 1 of the
performance verification test record as the absolute accuracy.
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...