212
Chapter 9
Performance Verification Tests
7. Reference Level Accuracy
7. Reference Level Accuracy
A 50 MHz CW signal is applied to the Input 50
Ω
of the analyzer
through two step attenuators. The amplitude of the source is decreased
in 10 dB steps and the analyzer marker functions are used to measure
the amplitude difference between steps. The source’s internal
attenuator is used as the reference standard. The test is performed in
both log and linear amplitude scales.
It is only necessary to test reference levels as low as
−
90 dBm (with
10 dB attenuation) since lower reference levels are a function of the
analyzer microprocessor manipulating the trace data. There is no error
associated with the trace data manipulation.
Equipment Required
Synthesized signal generator
1 dB step attenuator
10 dB step attenuator
Attenuator/ switch driver (if programmable step attenuators are
used)
Cable, Type-N 152-cm (60-in)
Cable, BNC 122-cm (48-in) (2 required)
Attenuator interconnect kit
Adapter, Type-N (m) to BNC (f) (2 required)
Additional Equipment for 75
Ω
Input
Pad, minimum loss
Adapter, Type-N (f), to BNC (m), 75
Ω
Procedure
Calculate Actual Attenuation Errors
1. From the calibration data supplied with the 10 dB step attenuator,
enter into column 2 of
Table 9-7
and
Table 9-8
the actual attenuation
for the corresponding nominal attenuation settings. If using a
programmable attenuator, enter the data for the section three 40 dB
step.
2. Calculate the reference attenuation error by subtracting 40 dB from
the actual attenuation for the 40 dB setting, and enter below.
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...