Chapter 3
59
Troubleshooting
Error Messages
CAUTION
Exposing the analyzer to high levels of input power over a prolonged
time period can damage the circuitry.
When operating remotely, the status bit should be monitored for this
condition to avoid incorrect measurement results.
PARAMETER ERROR: _ _ _
The specified parameter is not
recognized by the spectrum analyzer. See the appropriate
programming command description to determine the correct
parameters. (U)
Password required
Indicates that the service function cannot be
accessed without the password. (U)
Power-up selftest failed Boot ROM checksum:
Indicates a
test of the bootrom produced a checksum error. (H)
Power-up selftest failed RAM error:
Indicates a test of
RAM produced an error. (H)
Power-up selftest failed Checksum:
Indicates a test of the
instrument ROM produced a checksum error. (H)
Require 1 signal > PEAK EXCURSION above PEAK
THRESHOLD
Indicates that the N dB PTS routine cannot locate a
signal that is high enough to measure. The signal must be greater
than the peak excursion above the threshold level to measure. (U)
Require 3 signals > PEAK EXCURSION above PEAK
THRESHOLD
Indicates that the % AM routine cannot locate three
signals that are high enough to measure. The signals must be
greater than the peak excursion above the threshold level to
measure. (U)
Require 4 signals > PEAK EXCURSION above PEAK
THRESHOLD
Indicates that the TOI routine cannot locate four signals
that are high enough to measure. The signals must be greater than
the peak excursion above the threshold level to measure. (U)
Required Option Not Installed
Indicates that the function
selected requires optional hardware that is not currently available.
(U)
RF LO Unlocked
Indicates that the local oscillator within the
instrument’s RF assembly is unlocked. (H)
Attenuator
Setting
Attenuator
Coupling
Input Power Level
(50
Ω
)
Input Power Level
(75
Ω
)
≥
15 dB
Auto or Man
33 dBm
±
3 dB (nominal)
76 dBmV
±
3 dBmV
<15 dB
Man
13 dBm
±
7 dB (nominal)
68 dBmV
±
7 dBmV
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...