180
Chapter 8
Required Equipment
Test Equipment
Test Equipment
Periodically verifying operation
The analyzer requires periodic v
erification
of operation. Under most
conditions of use, you should test the analyzer at least once a year with
the complete set of performance verification tests.
Table 8-1
Recommended Test Equipment
Equipment
Critical Specifications for Equipment
Substitution
Recommended
Model
Use
a
Digital Multimeter
Input Resistance:
≥
10 megohms
Accuracy:
±
10 mV on 100 V range
HP 3458A
P,A,T
DVM Test Leads
For use with HP 3458A Digital Multimeter
HP 34118B
T
Universal Counter
Time Interval Range: 25 ms to 100 ms Single
Operation Range: +2.5 Vdc to
−
2.5 Vdc
HP 53132A or
HP 5316B
P,A,T
Frequency
Standard
Frequency: 10 MHz Timebase Accuracy
(Aging): < 1
×
10
-9
/day
HP 5071A
P,A
Measuring Receiver
Compatible with Power Sensors dB Relative
Mode Resolution: 0.01 dB
Reference Accuracy:
±
1.2%
HP 8902A
P,A,T
Microwave
Frequency Counter
Frequency Range: 9 MHz to 7 GHz
Timebase Accuracy (Aging): <5
×
10
-10
/day
HP 5352B
T
Oscilloscope
Bandwidth: dc to 100 MHz
Vertical Scale Factor of 0.5 V to 5 V/Div
HP 54501A
T
Power Meter
(Option 1DP)
Compatible with power sensors dB relative
mode.
Resolution: 0.01 dB
Reference accuracy:
±
1.2%
HP 436A
P,A,T
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...