90
Chapter 4
Front-Panel Key Reference
Measure
Amplitude
allows you to enter an amplitude value into a limit line
table. Key Access:
Measure
,
Limits
,
Edit Limits
,
Edit Line
Type
accesses the following
Type
menu keys used to select a
type of limit line:
Slope
The slope key draws a straight line between the
coordinate point of the current segment and the
coordinate point of the next segment, producing limit
line values for all frequencies between the two points.
Key Access:
Measure, Limits
,
Edit Limits
,
Edit Line
,
Type
Flat
The flat key draws a zero-slope line between the
coordinate point of the current segment and the
coordinate point of the next segment, producing limit
line values equal in amplitude for all frequencies
between the two points. If the amplitude values of the
two segments differ, the limit line “steps” to the
frequency value of the second segment. Key Access:
Measure, Limits
,
Edit Limits
,
Edit Line
,
Type
Point
The point key specifies a limit value for the
coordinate point. It specifies a limit at a single
frequency or time, and for no other frequency/time
points. For an upper limit line, a point segment is
indicated by a line drawn from the coordinate point,
vertically off the top of the screen. For a lower limit
line, a point segment is indicated by a line drawn from
the coordinate point vertically off the bottom of the
screen. The line will continue to the right edge of the
display, but testing stops at the defined point.
The point segment type should be used as the last
segment in the limit line table. However, if the last
segment in the table is not specified, an implicit point is
automatically used. If a point segment at the right edge
of the display is not desired, add a last-point segment to
the limit line table that is higher in frequency than the
displayed stop frequency. Key Access:
Measure, Limits
,
Edit Limits
,
Edit Line
,
Type
Del Segment
deletes the limit line entry for the selected segment
number. Front-Panel Key Access:
Measure
,
Limits
,
Edit Limits
,
Edit Line
Delete Limits
allows you to purge data from the limit line tables. Key
Access:
Measure
,
Limits
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...