Chapter 9
263
Performance Verification Tests
16. Other Input Related Spurious Responses
16. Other Input Related Spurious
Responses
A synthesized source and the spectrum analyzer are set to the same
frequency and the amplitude of the source is set to
−
10 dBm. A
marker-amplitude reference is set on the spectrum analyzer. The source
is then tuned to several different frequencies where image responses
could occur. At each source frequency, the source amplitude is set to
−
10 dBm and the amplitude of the response, if any, is measured using
the spectrum analyzer marker function. The marker-amplitude
difference is then compared to the specification.
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized sweeper
Measuring receiver (used as a power meter)
Power sensor, 100 kHz to 1500 MHz
Adapter, Type-N (f) to APC 3.5 (f)
Adapter, Type-N (f) to Type-N (f)
Cable, Type-N, 152-cm (60-in)
Additional Equipment for 75
Ω
Input
Power sensor, 75
Ω
Adapter, Type-N (f), to BNC (m), 75
Ω
Adapter, BNC (m), to BNC (m), 75
Ω
Pad, minimum loss
Procedure
1. Zero and calibrate the measuring receiver and 100 kHz to 1500 MHz
power sensor in log mode (power reads out in dBm), as described in
the measuring receiver operation manual. Enter the power sensor’s
500 MHz Cal Factor into the measuring receiver.
75
Ω
Input only: Use a 75
Ω
power sensor.
2. Press
INSTRUMENT PRESET
on the synthesized sweeper and set the
controls as follows:
CW
,
542.8 MHz
POWER LEVEL
,
−
10 dBm
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...