22. Tracking Generator Level Flatness, Options 1DN and 1DQ
10.Set the DVM to measure AC Volts. Press the following DVM keys so
that it reads out in dBm:
50 STORE 4
(sets resistance to 50
Ω
for dBm calculation)
MATH 4
(sets DVM to read out in dBm)
11.Set the spectrum analyzer center frequency to 9 kHz and press
Single Sweep
. Record the DVM readout in column 2 of
Table 9-25
.
12.Repeat
step 11
for all center frequencies listed in
Table 9-25
.
Table 9-25
Tracking Generator Level Flatness Worksheet, 100 kHz
13.Subtract the 100 kHz Level Flatness readout in
Table 9-24
from the
100 kHz DVM Readout in
Table 9-25
and record as the DVM Offset
at 100 kHz.
DVM Offset ____________ dB
For example, if the Level Flatness reading from
Table 9-24
is
+0.7 dB and the DVM Readout from
Table 9-25
is
−
0.53 dBm, the
DVM offset would be
−
1.23 dB.
14.Add the DVM Offset from
step 13
to each of the DVM Readouts in
Table 9-25
and record as the Corrected Level Flatness in column 3.
For example, if the DVM Readout from
Table 9-25
is +0.22 dBm, and
the DVM Offset is
−
1.23 dB, the corrected readout would be
−
1.01 dB.
15.Locate the most positive Level Flatness reading in
Table 9-24
and
Table 9-25
for frequencies of 10 MHz and less and record this value
as TR Entry 1 of the performance verification test record.
75
Ω
RF Out only: Do not use the data in
Table 9-25
.
Center
Frequency
DVM Readout
dBm
Corrected Level
Flatness dB
9 kHz
20 kHz
40 kHz
60 kHz
80 kHz
100 kHz
DVM
Level Flatness
–
DVM Offset
=
DVM
DVM Offset
+
Corrected Readout
=
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...