Chapter 9
237
Performance Verification Tests
11. Frequency Readout and Marker Count Accuracy
11. Frequency Readout and Marker Count
Accuracy
The frequency readout accuracy of the analyzer is tested with an input
signal of known frequency. By using the same frequency standard for
the analyzer and the synthesized sweeper, the frequency reference
error is eliminated.
There is no related adjustment for this performance test.
Equipment Required
Synthesized sweeper
Adapter, Type-N (f) to APC 3.5 (f)
Cable, Type-N, 152-cm (60-in)
Cable, BNC, 122-cm (48-in)
Additional Equipment for 75
Ω
Input
Pad, minimum loss
Adapter, Type-N (f), to BNC (m), 75
Ω
Procedure
This performance test consists of two parts:
Part 1: Frequency Readout Accuracy
Part 2: Marker Count Accuracy
Perform Part 1 before Part 2.
Part 1: Frequency Readout Accuracy
1. Connect the equipment as shown in
Figure 9-10
. Remember to
connect the 10 MHz REF OUT of the synthesized sweeper to the 10
MHz REF IN of the analyzer.
2. Perform the following steps to set up the equipment:
• Press
INSTRUMENT PRESET
on the synthesized sweeper, then set
the controls as follows:
CW
,
1.490 GHz
POWER LEVEL
,
−
10 dBm
• Press
Preset
on the analyzer, then wait for the preset routine to
finish. Set the analyzer by pressing the following keys:
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...