13. Frequency Response
7. Set the synthesized signal generator frequency to 100 kHz.
75
Ω
Input: Set frequency to 1 MHz.
8. Adjust the synthesized sweeper power level for a spectrum analyzer
marker amplitude reading of
−
14 dBm
±
0.10 dB.
9. Press
Frequency
,
100 kHz
on the spectrum analyzer.
75
Ω
Input: Press,
Frequency
,
1 MHz
10.Set the sensor Cal Factor on the measuring receiver, then record the
negative of the power ratio displayed on the measuring receiver in
column 2 of
Table 9-15
as the Error Relative to 50 MHz at 100 kHz.
75
Ω
Input: Error relative to 50 MHz at 1 MHz.
11.Set the synthesized signal generator CW to 100 MHz.
12.Press
FREQUENCY
,
100 MHz
on the spectrum analyzer.
13.Adjust the synthesized sweeper power level for a spectrum analyzer
marker amplitude reading of
−
14 dBm
±
0.05 dB.
14.Set the sensor Cal Factor on the measuring receiver, then record the
negative of the power ratio displayed on the measuring receiver in
column 2 of
Table 9-15
as the Error Relative to 50 MHz at 100 MHz.
15.On the synthesized signal generator, press
FREQUENCY
, and
↑
(step-up key), then on the spectrum analyzer, press
Frequency
, and
↑
(step-up key).
16.Repeat
step 13
and
step 14
for each new frequency, entering the
power sensor Cal Factor into the measuring receiver for each
frequency setting as indicated in
Table 9-15
.
17.This step applies only to analyzers with 75
Ω
inputs.
In
Table 9-15
, starting with the error at 1 MHz, subtract column 4
(System Error) from column 2 (Error Relative to 50 MHz), and
record the result in column 5 as the Corrected Error.
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...