Chapter 9
303
Performance Verification Tests
24. Non-Harmonic Spurious Outputs, Options 1DN and 1DQ
Figure 9-28
Non-Harmonic Spurious Outputs Test Setup
2. On the spectrum analyzer, press
Single Sweep
.
It is only necessary to perform the next step if more than 2 hours have
elapsed since a front-panel calibration of the microwave spectrum
analyzer has been performed.
The microwave spectrum analyzer should be allowed to warm up for
at least 30 minutes before proceeding.
3. Perform a front-panel calibration of the microwave spectrum
analyzer by performing the following steps:
Note that the following steps are for an HP 8566A/B microwave
spectrum analyzer, the steps may be different if you are using
another microwave spectrum analyzer.
• Connect a BNC cable between CAL Output and RF Input.
• Press
2
−
22 GHz
(
INSTR PRESET
),
RECALL
,
8
. Adjust AMPTD CAL
for a marker amplitude reading of
−
10 dBm.
• Press
RECALL
,
9
. Adjust FREQ ZERO for a maximum amplitude
response.
• Press
SHIFT
,
FREQUENCY SPAN
to start the 30 second internal
error correction routine.
• Press
SHIFT
,
START FREQ
to use the error correction factors just
calculated.
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...