306
Chapter 9
Performance Verification Tests
24. Non-Harmonic Spurious Outputs, Options 1DN and 1DQ
17.Calculate the difference between the amplitude of marked signal
and the fundamental amplitude as listed in
Table 9-27
.
For example, if the fundamental amplitude for a fundamental
frequency of 10 MHz is +1.2 dBm and the marker amplitude is
−
40.8 dBm, the difference is
−
42 dBc.
Record this difference as the non-harmonic response amplitude for the
appropriate spectrum analyzer center frequency and microwave
spectrum analyzer start and stop frequency settings in
Table 9-28
.
18.If a true non-harmonic spurious response is not found, record
“NOISE” as the non-harmonic response amplitude in
Table 9-28
for
the appropriate spectrum analyzer center frequency and microwave
spectrum analyzer start and stop frequency settings.
19.Repeat
step 13
through
step 18
for the remaining microwave
spectrum analyzer settings for start frequency, stop frequency, and
resolution bandwidth; and for the spectrum analyzer center
frequency setting of 10 MHz.
20.Repeat
step 12
through
step 19
with the spectrum analyzer center
frequency set to 750 MHz.
21.Repeat
step 12
through
step 19
with the spectrum analyzer center
frequency set to 1.5 GHz.
22.Locate in
Table 9-28
the most-positive non-harmonic response
amplitude. Record this amplitude as the highest non-harmonic
response amplitude in TR Entry 1 of the performance verification
test record.
Table 9-28
Non-Harmonic Responses Worksheet
a. 75
Ω
RF outputs: Omit this frequency range.
b. 75
Ω
RF outputs: Set the start frequency to 1 MHz.
Non-Harmonic Amplitude
Marker Amplitude
Fundamental Amplitude
–
=
Microwave Spectrum
Analyzer Settings
Non-Harmonic Response Amplitude (dBc)
Start
Freq
(MHz)
Stop
Freq
(MHz)
Resolution
Bandwidth
at 10 MHz
Center Freq
at 750 MHz
Center Freq
at 1.5 GHz
Center Freq
9 kHz
a
0.1
a
300 Hz
a
0.1
b
5.0
10 kHz
5.0
55
100 kHz
55
1240
1 MHz
1240
1500
1 MHz
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...