Chapter 9
243
Performance Verification Tests
13. Frequency Response
13. Frequency Response
The output of the synthesized signal generator is fed through a power
splitter to a power sensor and the spectrum analyzer. The synthesized
signal generator’s power level is adjusted at 50 MHz to place the
displayed signal at the spectrum analyzer center horizontal graticule
line. The measuring receiver, used as a power meter, is placed in RATIO
mode. At each new synthesized signal generator frequency and
spectrum analyzer center frequency setting, the synthesized signal
generator’s power level is adjusted to place the signal at the center
horizontal graticule line. The measuring receiver displays the inverse of
the frequency response relative to 50 MHz.
For frequencies below 100 kHz, a synthesizer/function generator is used
in place of the synthesized signal generator. A DVM and 50
Ω
load are
used in place of the power sensor and measuring receiver.
The related adjustments for this performance test are “Corrected
Flatness” and “Uncorrected Flatness.”
Analyzers with 75
Ω
inputs are tested only down to 1 MHz. Also, a
system flatness characterization is performed.
Equipment Required
Synthesized signal generator
Measuring receiver (used as a power meter)
Synthesizer/function generator
Power sensor, 100 kHz to 1800 MHz
Power splitter
Adapter, Type-N (f) to APC 3.5 (f)
Adapter, Type-N (m) to Type-N (m)
Adapter, Type-N (m) to BNC (f) (2 required)
Dual banana plug to BNC (f)
BNC tee (BNC f,m,f)
Cable, BNC, 122-cm (48-in) (2 required)
Cable, Type-N, 152-cm (60-in)
50
Ω
termination BNC (m)
DVM HP3458, or HP34401A only
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...