Chapter 9
217
Performance Verification Tests
7. Reference Level Accuracy
Peak Search
Marker
,
Marker
∆
12.Set the 10 dB step attenuator and analyzer reference level according
to
Table 9-8
. At each setting, do the following:
a. Press
Single Sweep
on the analyzer.
b. Press
Peak Search
c. Record the marker delta amplitude reading in column 5 of
Table 9-8
.
d. Add the actual attenuation error to the analyzer marker
∆
amplitude and enter the result in the performance test record.
NOTE
When entering reference levels on analyzers with 75
Ω
inputs, enter the
values in dBmV with 0.01 dB precision. That is,
−
8.24 dBmV not
−
8.2 dBmV. The reference level annotation rounds to the nearest 0.1 dB,
but internal values are to the nearest 0.01 dB.
Table 9-8
Reference Level Accuracy, Linear Mode
a. Use the dBm column values for analyzers with a 50
Ω
input and the dBmV column for analyzers with a 75
Ω ινπυτ.
13.Press
Preset, System, Alignments, Auto Align, All
.
10 dB
Attenuator
Nominal
Attenuation
dB
10 dB
Attenuator
Actual
Attenuation
dB
Attenuation
Error dB
Analyzer
Reference Level
a
dBm
dBmV
Analyzer
Marker
∆
Amplitude
dB
TR Entry
30
0 (Ref)
−
30
+21.76
0 (Ref)
20
−
20
+31.76
9
10
−
10
+41.76
10
40
−
40
+11.76
11
50
−
50
+1.76
12
60
−
60
−
8.24
13
70
−
70
−1
8.24
14
80
−
80
−2
8.24
15
90
−
90
−3
8.24
16
Содержание ESA-L1500A
Страница 12: ...11 1 Preparing for Use and Printing ...
Страница 28: ...27 2 Getting Started ...
Страница 44: ...Chapter 2 43 Getting Started Making a Measurement Figure 2 5 Reading the Amplitude and Frequency ...
Страница 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Страница 48: ...47 3 Troubleshooting ...
Страница 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Страница 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Страница 118: ...117 5 Menu Maps ...
Страница 119: ...118 Chapter5 Menu Maps ...
Страница 120: ...Chapter 5 119 Menu Maps ...
Страница 121: ...120 Chapter5 Menu Maps ...
Страница 122: ...Chapter 5 121 Menu Maps ...
Страница 124: ...Chapter 5 123 Menu Maps ...
Страница 125: ...124 Chapter5 Menu Maps ...
Страница 127: ...126 Chapter5 Menu Maps a Frequency is replaced with Time when the X Axis Units Freq Time selection is Time ...
Страница 128: ...Chapter 5 127 Menu Maps ...
Страница 129: ...128 Chapter5 Menu Maps ...
Страница 130: ...Chapter 5 129 Menu Maps ...
Страница 131: ...130 Chapter5 Menu Maps ...
Страница 135: ...134 Chapter5 Menu Maps ...
Страница 136: ...135 6 Options and Accessories ...
Страница 145: ...144 Chapter6 Options and Accessories Accessories ...
Страница 146: ...145 7 Specifications ...
Страница 179: ...178 Chapter7 Specifications Regulatory Information ...
Страница 180: ...179 8 Required Equipment ...
Страница 187: ...186 Chapter8 Required Equipment Test Equipment ...
Страница 308: ...307 10 Performance Verification Test Record ...
Страница 321: ...320 Chapter10 Performance Verification Test Record HP ESA L1500A Performance Verification Test Record ...