Model 3585A
Adjustments
5·37. Range Up Detector Adjustment
a. Set the synthesizer controls for:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30kHz
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . -24dBm
b. Set the 3585A controls for:
RECALL 601
INSTRUMENT PRESET
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -25dBm
AUTO RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
c. Adjust the RANGE UP THRESHOLD, AIRI73, so that the front panel OVERLOAD
light is lit. Now adjust AIR173 so that the OVERLOAD light just goes out.
5·38. Range Down Detector Adjustment
a. Set the 3585A to the OdBm RANGE.
b. Set the synthesizer for an AMPLITUDE of -6dBm.
c. Connect a dc voltmeter (lOV range) to the Digital motherboard, pin A45B 19 or
A40TP2 ('L' RNGD). This is accessable from the bottom of the instrument as shown in
Figure 5-4 1 .
A 7 6
@ @
o GND
TI'3
•
I.ftNGU
,p,
0
L".D
TI'1 o LTRIP
A 4 0
358511-5-6
Figure 5·4 1 . Range Down Monitor Point (LRNGD)
5-63