Adjustments
Model 35 85A
5-20
5· 1 3. Fractional
N
Adjustments
a. Connect a DVM set for DC volts to A34TP5 . Adjust A34R32 for 5 . 3Vdc
±
0.05V.
b . Verify that A34TP6 measures + 1 5 . 0Vdc
±
0.8V and that A34TP8 measures
- 1 5 . 0Vdc
±
0 . 8V .
c. Disconnect the DVM .
d. Turn the 3585A power off. Place the A3 1 board on a PC board extender. Turn the
3585A power on.
e. Set the 35 85A controls for:
RECALL 601
INSTRUMENT PRESET
RES . BW .
. . . . .
. .
.
. .
. . . .
.
. .
. .
.
.
.
.
. .
. .
.
. .
. .
.
.
3KHz
RES . BW HOLD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
START FREQUENCY . . . . . . . . . . . . . . . . . . . . . 0.4MHz
STOP FREQUENCy . . . . . . . . . . . . . . . . . . . . . 1 . 65MHz
MANUAL ENTRy . . . . . . . . . . . . . . . . . . . . . . . . 0.4MHz
f. Connect the DVM to A3 1TP l and adjust A3 1 L3 for + 7 .70Vdc
±
0.05V.
5 . 3Vdc
±
O . 0 5 V
R 3 2
+
1 5 . 0Vdc
±
O. 8V