Adjustments
Model 3585A
5-52
o. Check that the instrument down ranges to the -25dBm Range with no input signal.
p. Press the INSTRUMENT PRESET key of the 3585A.
q. Connect an Ohmmeter to the 500 input. Ground lead to the outer shell of the 500 in
put connector and the ohms lead to the cent er pin of the input connector.
r. The Ohmmeter should now read 500 ± 20.
s. Press the I MO Impedance key. This action terminat(�s the input with a 500 load.
t. The Ohmmeter should now read 500 ± 20. (This reading should be slightly different
than the previous 500 reading.)
u. Press the 750 Impedance key on the 3585�.
v. The Ohmmeter should now read 750 ± 20.
w. Press the I MO Impedance key. This action terminates the input with a 750 load.
x.
The Ohmmeter should now read 750 ± 20. (This reading should be slightly different
then the previous 750 reading.)
y. Set the 3585A controls for:
RECALL 601
INSTRUMENT PRESET
MANUAL ENTRy . . . . . . . . . . . . . . . . . . . . . . . . . 9MHz
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IdB
SAVE 1
z. Set the synthesizer controls for:
FREQUENCy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9MHz
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OdBm
aa. Connect the synthesizer output to the 3585A 500 input.
NOTE
Use a non-metalic adjusting tool for all Conversion section ad
justments.
NOTE
When more than one component is called out for adjustment in
any given step, adjust them in the order listed.
NOTE
Figure 5-45 shows the location of the Input/Conversion section
adjustments.