Model 3585A
Backdating
h . Adjust A l 3C39 for a maximum marker amplitude. If necessary, set the REf LEVEL so
that the marker remains within the graticule area.
i. Press the STOR A - B key.
j . Disconnect the synthesizer .
k . Connect the Tracking Generator output through a 500 termination to Al l J 1 .
1.
Set the -hp- 3585A to :
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50kHz
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300Hz
SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . cont
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10dB
B TRACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
m . Move the marker to the peak of the trace and press MKR - CF.
n . Adj ust A l 3C41 so that the trace is symmetrical about the marker .
Figure 7- 1 . Symmetry Adjustment, A 1 3C4 1
o . Narrow the FREQUENCY SPAN to 1 0Hz by using the STEP keys . As you narrow the
span, the peak of the trace will move to the left or right . When this occurs, move the marker
to the peak of the trace and press the MKR CF key.
p . Set the -hp- 3585A to :
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I dB
SWEEP TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 . 6 sec
B TRACE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
q. Move the marker to the peak of the trace and press MKR - CF.
r. Adj ust the Tracking Generator amplitude control so that the peak of the A trace and
peak of the B trace are equal in amplitude .
Scans by ARTEK MEDIA
=>
7-3