Performance Tests
Model 3585A
4-8
e. Place the marker on the most positive point of the CRT trace. The marker reading in
the upper right of the CRT will assist you in finding this point.
f. Enter the Offset and Marker reading on the Operational Verification Test Card.
g. Repeat Steps e and g for each of the Resolution Bandwidths listed in the Operational
Verification Test Card.
4· 1 2. Range Calibration
This test verifies that the Range Calibration system is working as specified.
Specification:
± 0.7dB (Equals the Reference Level Accuracy
( ± O.4dB) for a -25dBm signal plus the Amplitude
Linearity spec (0.3dB) for a signal 5dB below the
Reference Level)
Equipment Required:
Frequency Synthesizer . . . . . . . . . . . . . . . -hp- 3335A
Procedure:
a. Set the 3585A controls for:
INSTRUMENT PRESET
CENTER FREQUENCY . . . . . . . . . . . . . . . 1 50 kHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . -20 dBm
1 dB/DIV
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Hz
MANUAL SWEEP . . . . . . . . . . . . . . . . . . . . . . . . on
REF LEVEL TRACK . . . . . . . . . . . . . . . . . . . . . . off
b . Set the synthesizer controls for:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . 1 50 kHz
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . -25 dBm
c. Connect the 500 output of the synthesizer to the 500 input of the 3585A.
d. Set the 3585A controls for:
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
ENTER OFFSET
RANGE UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
e. Check the marker reading for all ranges, it should be less than ± 0.7 dB to verify that
the RANGE selected is within specification. Enter Marker readings on the Operational
Verification Test Card.