Model 3585A
Adjustments
k. Set the 3585A controls for:
RECALL 601
INSTRUMENT PRESET
RANGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . + 30dBm
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . + 27dBm
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . l dB
MANUAL ENTRY . . . . . . . . . . . . . . . . . . . . . . . . 350KHz
1.
Adjust A1 5R7 for a marker amplitude reading of + 27.00dBm.
m. Replace the metal cover on the A14 thru A16 boards and tighten down the associated
screws.
5·20. I.F. Filter Adjustments
NOTE
Use a non-metallic adjusting tool jor all I.F. Filter adjustments.
a. Turn the 3585A power off. Remove the aluminum cover on the A17, A18 and A 1 9
boards.
b. Place the A19 board on a PC extender. Restore power to the 3585A.
c. Move the test jumper A191 1 to the "T" position.
NOTE
The component locatorsjor the IF boards (A 1 7-A 19) are contain
ed on Figure 5-34.
d. Set the synthesizer for a Frequency of 350KHz and an Amplitude of -2.0dBm.
e. Remove the cable from the A1711 connector.
f. Set the 3585A controls for:
RECALL 609
INSTRUMENT PRESET
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . 350KHz
CF STEP SIZE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 .3Hz
RES . BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
[£Jl [[f8]
RES. BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Hz
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I dB
MANUAL SWEEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
CLEAR A
g. Connect the output of the synthesizer to a 50n termination. Connect the output of the
termination to the A1711 connector. Adjust synthesizer output level until it is
==
I dB below
3585A Reference Level.
5-35