Performance Tests
Model 3585A
4-46
4·39. Semi·Automatic Performance Test Procedure
a. Turn the calculator power off.
b . Insert the calculator ROM 's described in the Recommended Test Equipment list, Table
4- 1 , and into the slots under the calculator keyboard (see Figure 4-4) .
c . Check the rotary switch setting (Figure 4- 1 9) on the HP-IB interface cable. The pointer
should be on "7" . If the pointer is at some other setting, use a small screwdriver to set it on
"7" .
d . Turn the calculator, printer, synthesizer and spectrum analyzer power on .
e. Load the supplied performance test cassette tape (Part Number 03585- 1 000 1 ) into the
cassette tape slot .
f. Press
8
G)
on the calculator keyboard .
g. Press
h . After the run light has gone out, press
8
on the calculator keyboard .
i. From this point on, the calculator will give instructions for what to do next . The
manual does contain equipment set-ups for some of the tests . As these tests are encountered,
the calculator will refer you to the equipment set-up diagram in the manual .
Messages
Slots for ROM ' s
Figure 4· 1 9. HP 982SA Calculator
Sw itch
REAR OF
CALCULATOR