![HP 3585A Скачать руководство пользователя страница 115](http://html.mh-extra.com/html/hp/3585a/3585a_service-manual_161430115.webp)
5- 1 . INTRODUCTION
SECTION V
ADJUSTMENTS
This section contains complete adjustment procedures for the Model 3585A Spectrum
Analyzer. Table 5-1 lists the adjustments and their affected components. Figures 5-45 and
5-46 are foldout drawings found at the end of this section. These drawings show adjustment
locations throughout the instrument.
Table 5- 1 . Adjustment Locations
Paragraph
Adjustment
Location
Affected Components
Service Group
Low Voltage Power Supplies
5-6
A 7 5 R9 , R 1 5 , A 7 2 R3 1 ,
I
(A7 1 - 7 5 )
R 1 9
90MHz Reference 80ard (A2 1 )
5-7
A2 1 R 1 2 5
8- 1
1 OMHz Oven Oscillator (A8 1 )
5-8
A8 1 R2
8 - 1
Oven Output Shutdown
5-9
A8 1 R9
8 - 1
CRT Control and High Voltage
5- 1 0
A 6 7 R46,R6,R 1 0 5, R 1 1 6 ,
D- 2,4
Power Supply (A6 3 , 6 5 , 6 7 )
A 6 5 R 1 3 , A 6 3 R 4 , R 1 -
6 , R 3 8
CRT G raphics (A64 , 6 7 )
5 - 1 1
A 6 7 R 5 9 , R 8 5 , R 5 4 , R 8 0 ,
D-4
R3, R2, R 1
CRT Alphanumerics (A64)
5- 1 2
A 6 4 R 7 2 , C 2 3 , R4 8 , R 6 2 ,
D-3
R 1 4, R 1 6 , R 1
Fractional N (A3 1 -34)
5 - 1 3
A34R32,A3 1 L3,
8-4 , 5
A32R49 , R 5 6
L . O . Step Loop (A23-26)
5- 1 4
A 2 3 L 1
8-3
First L . O . VTO and Sum Loop
5- 1 5
A 2 7 R 2 , R 1 1
8-2
Video Filter and AID (A 1 5, 1 6 )
5- 1 6
A 1 6 R2 1 , R 1 9 ,A 1 5R4
A- 5 , 6
Log A m p a n d 30kHz Filter (A 1 4)
5- 1 7
A 1 4L 5 , L 7 , R 5 7 , R 5 3
A-3 , 4 , 5
A 1 7 R 1 05,A 1 5 R 7
Log A m p Slope Adjustment (A 1 4)
5- 1 8
A 1 4 R 4 3 , R 1 7 ,R 7 , R 8 , A-4
R 1 4 , R 2 1 , R 2 6
Reference Level DC Offset (A 1 5)
5- 1 9
A 1 5 R9,R7
A- 5
----
5-1