Adjustments
Model 3585A
5- 1 6
m . Move test jumper A64J l to the "N" position .
n. Adjust A64R62 (LD gain) for an overshoot condition (see Figure 5-1 3).
o . Adjust A64R62 so that the overshoot condition just disappears .
p . Adjust A67R59
(X
position) and A67R54
(X
gain) so that the ends of the retrace line
and bottom pattern line are aligned with the vertical lines of the CRT graticule (see Figure
5 - 1 2).
q. Adjust A67R85 (Y position) and A67R80 (Y gain) so that the CRT graticule lines cut
through the upper and lower lines of the displayed pattern (see Figure 5- 1 2) .
r. Replace the A67 board . Replace the screws that hold the board t o the chassis and the
protective plastic cover over the board.
N o Overshoot
Of Vertical Line
At This Point
R e t r a c e L i n e
A n d B o t t o m
Line Should Be
Straight
Corners Do Nbt
O v e r s h o o t
Graticule
Properly Adjusted CRT Display.
F o r
L i n e
Vertical Alignment With Graticule
Figure 5· 1 2. CRT Test Pattern
Last Right H a nd
C h a r a c t e r
Same Dista nce
From Graticule
Edge As Left
Character
Space of
1 /4
C h a r a c t e r
H e i g ht Above
Retrace Line
Retrace Line
C o r n e r D o e s
Not overshoot
Graticule
Space Of
1 /4
C h a r a c t e r
Height