o. Set the controls of synthesizer #2 for:
FREQUENCY . . . . . . . . . . . . . . . . . . . 33.0001 MHz
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . -25 dBm
p. Set the 3585A controls for:
CENTER FREQUENCY . . . . . . . . . 33.00005 MHz
CONT. SWEEP
q. Adjust attenuators for a - 3.1 dBm level on the 3585A.
r. Repeat Steps e thru m using the values in parenthesis.
s. This completes the test, disconnect all inputs to the 3585A.
4·27. Bandwidth Tests
These tests will verify that the 3585A meets its 3 dB, Bandwidth and Shape Factor specifica
tions.
Specification:
Resolution Bandwidth Accuracy
3 dB Bandwidth . . . . . . . . . ± 20% of BW setting at the 3dB points
Selectivity (Shape Factor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
<
1 1 : 1
Procedure:
a. Set the 3585A controls for:
RECALL 602
INSTRUMENT PRESET
CENTER FREQUENCY . . . . . . . . . . . . . . . 10 MHz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . 10 Hz
REFERENCE LEVEL . . . . . . . . . . . . . . . -24.5 dBm
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 dB.
RES. BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Hz
RES. BW HOLD . . . . . . . . . . . . . . . . . . . . . . . . . . on
h.
Initially this test checks the 3 dB points of each Resolution Bandwidth; therefore,
ignore the values in parenthesis until instructed otherwise.
c. Allow one complete sweep to occur. Now put the marker on the most positive point of
the trace, using the marker amplitude reading as your guide.
d. Set the 3585A controls for:
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
ENTER OFFSET