2
OPERATIONAL VERIFICATION TEST CARD (Cont'd)
AMPLITUDE LINEARITY TEST
(E)
(A)
(B)
(C)
(D)
3585A
Variable
Correction
Ideal
Correct
Marker
Attenuator
Factor·
Reading
Reading
Reading··
o
dB
-
-
-
-
00.0 dB
00.0 dB
00.0 dB
- 1 0 dB
- 1 0 . 0 dB
__
dB
dB
-20 dB
-20.0 dB
__
dB
d B
- 3 0 dB
-30.0 dB
__
dB
dB
-40 dB
-40.0 dB
__
dB
dB
- 50 dB
- 5 0 . 0 dB
__
dB
d B
- 6 0 d B
- 6 0 . 0 dB
__
dB
dB
-70 dB
-70.0 dB
__
dB
dB
-80 dB
-80 . 0 dB
__
dB
dB
-90 dB
-90.0 d B
__
dB
dB
* Correction factor must be obtained from attenuator calibration data.
* * If noise jitter is present, use average marker reading.
(A)
Synthesizer
Level
+ 1 0 dBm
o
dBm
- 1 0 dBm
- 20 dBm
- 30 dBm
- 40 dBm
- 50 dBm
- 60 dBm
- 70 dBm
- 80 dBm
REFERENCE LEVEL ACCURACY TESTS
(B)
(C)
(D)
3585A
3585A
Synthesizer Level
Reference
Marker
Minus The 3585A
Level
Reading
Marker Reading
+
1 0 dBm
dBm
o
dBm
dBm
- 1 0 dBm
dBm
- 20 dBm
dBm
- 30 dBm
dBm
- 40 dBm
dBm
- 50 dBm
dBm
- 60 dBm
dBm
- 70 dBm
dBm
- 80 dBm
dBm
50n FREQUENCY RESPONSE TEST
Test Limit ± . 5 dB
Range
- 2 5 dBm
- 20 dBm
- 1 5 dBm
- 1 0 dBm
- 5 dBm
o
dBm
5 dBm
1 0 dBm
1 5 dBm
20 dBm
25 dBm
30 dBm
Maximum Amplitude
Deviation
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
dB
d B
dB
dB
dB
(F)
Test
Tolerance
- -
- -
± 0 . 3 dB
± 0. 3 d B
± 0. 6 dB
± 0. 6 dB
± 1 . 0 dB
± 1 . 0 dB
± 1 . 0 dB
± 2 . 0 dB
± 2.0 dB
(E)
Test
Tolerance
± 0 . 4 dB
± 0 . 4 dB
± 0. 4 d B
± 0. 4 dB
± 0. 4 dB
± 0. 4 d B
± 0. 7 dB
± 0. 7 dB
± 1 . 5 dB
± 1 . 5 dB