Model 3585A
Adjustments
e. Allow a complete sweep to occur, then enter:
ENTER OFFSET
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 KHz
f. Adjust A1 7R20 for a marker amplitude reading of .OOdB.
5·30. First Le Stage Adjustment
a. Move test jumper A17 J2 to the "OP" position and test jumper A 1713 to the "T" posi
tion.
b. Set the 3585A controls for:
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . OFF
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . 3 .3KHz
dB/DIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IdB
A-B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on
c. Adjust A1 7L4 so that the A trace approximates a straight, horizontal line.
d. Set the 3585A controls for:
A-B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . off
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30KHz
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
' . . . . . . . . . on
e. Allow a complete sweep to occur, then enter:
ENTER OFFSET
RES BW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 KHz
f. Adjust A17R12 for a marker amplitude reading of .OOdB.
g. Move test jumper A1 713 to the "OP" position. Check that all test jumpers on the A17
board are in the "OP " position.
h. Turn the 3585A power off. Remove the A17 board from the PC extender and replace it
in the card nest.
i. Replace the metal cover over the A17 - A19 boards. Insert and tighten down all screws
that hold down the cover. Restore power to the 3585A.
NOTE
If continuing DO NOT reconnect green cable at this time.
5-45