![Honeywell 4022 Скачать руководство пользователя страница 129](http://html.mh-extra.com/html/honeywell/4022/4022_system-manual_152613129.webp)
Cable End Electrical Assemblies and Precision Measurement Processor
Cable End electrical overview
p/n 6510020208
10/18/07
6-3
resides in a PMP (4022-31 and 4022-40) or in the scanner itself where there
is no PMP (4022-30).
The Encoder Assembly sends signals relating to scanner position and
velocity back to the Quality Control System or PMP.
The scanner cable end also houses a 56 VAC Heater Transformer for
supplying power to the resistive heaters located inside the heads.
The Termination Board Assembly consists of the Termination Board and
the Logic Daughter Board plugging into the Termination Board. These
boards are discussed in detail in Section 6.3.
Some of these assemblies are shown in Figure 6-1.