96
Gradual Voltage Rise Test
Setting the voltage holding time for each step
The voltage holding time for each step is set.
1
1
1
Press
[BASIC]
>
[HOLD TIME]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
20 s
*
1
,
60 s
,
300 s
,
TRIG
, any time
Setting range:
1 s
to
300 s
*1. Default setting
When
[TRIG]
is set, TRIG operation is performed. Press
the
START
button to proceed to the next step.
Vs:
Initial voltage
Ve:
End voltage
Vbd: Dielectric breakdown voltage
∆V:
Voltage steps
Dielectric breakdown
Ve
Vbd
Vs
∆V
HOLD
START
START
START
START
START
START
HIOKI ST5680A961-00