174
Timing Charts
Timing chart for the program test mode
Contact check when pass judgment in the program mode
STAR
T
READ
Y
W
-M
O
D
E
STEP
_EN
D
PR
O
G
_E
N
D
IR
-M
O
D
E
HV
_O
N
PA
SS
(T
EST
)
(P
R
O
BE)
(R
ISE
)
(M
EA
S)
TE
ST
15 ms typ.
DCW
Start voltage
Output voltage
5 ms min.
Test voltage
Interval
(pass, trigger)
Interval (pass)
(total judgement) Pass
Interval (pass)
DANGER lamp lit
Test time
Rise time
0.1 s
Test time
Interval time
(trigger)
Rise time
Interval
time
Interval
time
Contact check period
Contact check period
Contact check period
IR
DCW
Discharge time
Discharge time
Dielectric breakdown
5 ms min.
10 ms min.
3 ms typ.
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...