209
Functional Specifications
Output voltage monitoring
function
Operations
The output is shut down and the test ends when the output
voltage exceeds
±
(5% of the s10 V) after the voltage
rise time or the voltage deviates from the test voltage during
the test and does not fall within
±
10 dgt of the set voltage
within 5 seconds.
Minimum resistance value
display
Operations
The minimum value of the measured insulation resistance
value during a test can be displayed.
Insulation resistance test
end mode
Operations
Sets the conditions to end an insulation resistance test.
Operation mode
Default setting
Time:
Pass:
Fail:
Time
Performs the test for the set duration and makes
a judgment with the value at the end of the test.
Ends the test at the time of pass judgment within
the set duration (excluding the delay time).
Ends the test at the time of fail judgment within
the set duration (excluding the delay time).
-3. W-IR/IR-W test mode
Operations
W-IR mode Performs a withstand voltage test and then an insulation resistance
test.
IR-W mode Performs an insulation resistance test and then a withstand voltage
test.
Setting details
Withstand voltage
test
Insulation resistance
test
Interval time
Operations for fail
judgment
Test conditions set in the withstand voltage test mode
Test conditions set in the insulation resistance test mode
0.1 s to 100.0 s, TRIG
When TRIG is set, pressing the
START
button proceeds to
the next step.
Default setting: 0.1 s
Test end, test continue
Default setting: Test end
-4. Program mode
Operations
Operations are performed automatically with DCW and IR combined.
Maximum number of steps
50
Step setting items
The test conditions similar to those for a single test can be set.
(The test time cannot be set to off.)
Program setting items
Step interval time
Operations for fail
judgment
0.1 s to 100.0 s, TRIG
Default setting: 0.1 s
When TRIG is set, pressing the
START
button proceeds to
the next step.
Test end, test continue
Default setting: Test end
Specifications
HIOKI ST5680A961-00
Содержание ST5680
Страница 2: ...HIOKI ST5680A961 00 ...
Страница 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Страница 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Страница 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Страница 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Страница 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Страница 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Страница 250: ...250 HIOKI ST5680A961 00 ...
Страница 253: ...www hioki com HIOKI ST5680A961 00 ...