145
Self-test Function
Calibrating the touch screen
You can calibrate the touch screen position.
1
1
1
Press
[SYS]
>
[TEST]
>
[EXEC]
of
[TOUCH
SCREEN CAL]
on the measurement screen.
2
2
2
2
2
Press and hold down the center of the two
[+]
symbols for approximately for 2 seconds.
Green circles appear. When the two
[+]
symbols change
to green circles,
[SET]
is displayed.
3
3
3
Press
[SET]
.
IMPORTANT
If
[SET]
is not displayed, repairs are necessary. Please contact your authorized Hioki distributor
or reseller.
System Settings
HIOKI ST5680A961-00